9801404075 nsbmet@gmail.com

CT single Syringe Injection System

  • Rs



Technical specification Description 
Electrical requirement 220VAC/ 110VAC, 50 or 60Hz, 200VA
Operation Interface     Available in English, French, Spanish, German and Chinses    
Flow Rate 0.1-10 m/sec. in increments of 0.1 ml/sec
Syringe Capacity  100ml for Zenith- C10; 200ml for Zenith-C20
Volume 0.1 ml to syringe capacity in 0.1 ml increments 
Pressure 

Preset limit:300 psi default, user settable from 50to 350 psi in 1 psi increments 

Display: Real- time pressure curve 

Change threshold: Stop injection if pressure changing suddenly 

Alarm: Voice alarm and stop injection when over preset limit 

Injection or scan Delay 0 to 3559 seconds in 1 sec. increments 
AutofillSettable flow rate from 3- 8ml/sec., volume from 10ml to syringe capability 
Multi- Phase  1-8 phases per injection 
Pause Phase 0~ 999 sec. in 1 sec. increments 
Hold Capability more then 30 minutes 
Storage Capability 1000 protocols 
Expel air locked Unable to inject unless expelling air 
Test injection Settable flow rate from 0.1- 5ml/sec., volume from 0.1ml to 10ml 
Heat maintainer 35~38°C (optional)
Hand switch Hand switch control pause and continue injection (Optional) 
Emergency stop function 

Clicking anywhere on the touch screen of the console or pressing any button on the

injection arm to stop protocol injection  

Dimension 

Injector Arm and Pedestal: L: 22.8* (58.0cm), W: 12.5 (20.1cm), H: 50.2* (127.5cm)

Console: L: 4.55* (11.55 cm). W: 12.68* (31.8cm), H: 9.77* (24.81cm )

Main Controller: L:11.9* (30.2cm) W: 7.4* (9.1cm) H: 4.7* (12cm)

Weight 

Injector Arm and pedestal: 16.5 kg

Console: 4.1 kg 

Main Controller: 6.2kg

User-friendly Operations 

  • Programming complex protocols easily on 12" wide touch screen of console 
  • Performing all operations conveniently either on the console or on the injector arm
  • Smart buttons of different functions making operation simply 

Reliability & Safety 

  • High precision of flow rate and volume of injection 
  • Comprehensive real-time monitoring the pressure and system operating states
  • Air expelling lock and warning before injections to reduces risk of air embolism

Science & Humanized Design 

  • The injection arm's height and direction are adjustable 
  • The pedestal can be moved and locked randomly at expected place.
  • The accessary tray frees the user's hands to focus on the operation.


Multi- purpose functionality 

  • Designed for use in multiple imaging environment, including CT, peripheral vascular surgery, Mammography, HSG, IV Urography and more


Share



Technical specification Description 
Electrical requirement 220VAC/ 110VAC, 50 or 60Hz, 200VA
Operation Interface     Available in English, French, Spanish, German and Chinses    
Flow Rate 0.1-10 m/sec. in increments of 0.1 ml/sec
Syringe Capacity  100ml for Zenith- C10; 200ml for Zenith-C20
Volume 0.1 ml to syringe capacity in 0.1 ml increments 
Pressure 

Preset limit:300 psi default, user settable from 50to 350 psi in 1 psi increments 

Display: Real- time pressure curve 

Change threshold: Stop injection if pressure changing suddenly 

Alarm: Voice alarm and stop injection when over preset limit 

Injection or scan Delay 0 to 3559 seconds in 1 sec. increments 
AutofillSettable flow rate from 3- 8ml/sec., volume from 10ml to syringe capability 
Multi- Phase  1-8 phases per injection 
Pause Phase 0~ 999 sec. in 1 sec. increments 
Hold Capability more then 30 minutes 
Storage Capability 1000 protocols 
Expel air locked Unable to inject unless expelling air 
Test injection Settable flow rate from 0.1- 5ml/sec., volume from 0.1ml to 10ml 
Heat maintainer 35~38°C (optional)
Hand switch Hand switch control pause and continue injection (Optional) 
Emergency stop function 

Clicking anywhere on the touch screen of the console or pressing any button on the

injection arm to stop protocol injection  

Dimension 

Injector Arm and Pedestal: L: 22.8* (58.0cm), W: 12.5 (20.1cm), H: 50.2* (127.5cm)

Console: L: 4.55* (11.55 cm). W: 12.68* (31.8cm), H: 9.77* (24.81cm )

Main Controller: L:11.9* (30.2cm) W: 7.4* (9.1cm) H: 4.7* (12cm)

Weight 

Injector Arm and pedestal: 16.5 kg

Console: 4.1 kg 

Main Controller: 6.2kg

User-friendly Operations 

  • Programming complex protocols easily on 12" wide touch screen of console 
  • Performing all operations conveniently either on the console or on the injector arm
  • Smart buttons of different functions making operation simply 

Reliability & Safety 

  • High precision of flow rate and volume of injection 
  • Comprehensive real-time monitoring the pressure and system operating states
  • Air expelling lock and warning before injections to reduces risk of air embolism

Science & Humanized Design 

  • The injection arm's height and direction are adjustable 
  • The pedestal can be moved and locked randomly at expected place.
  • The accessary tray frees the user's hands to focus on the operation.


Multi- purpose functionality 

  • Designed for use in multiple imaging environment, including CT, peripheral vascular surgery, Mammography, HSG, IV Urography and more


Add a Review

Your rating: