Dimensions and Weight | Unit dimensions (overall) | 66 cm x 38 cm x 49.5 cm |
Unit dimensions (top light) | 51.6 cm x 12.5 cm x 6.3 cm | |
Unit dimensions (bottom light) | 64.7 cm x 38 cm x 10 cm | |
Shipping (boxed) dimensions | 45 cm x 73 cm x 55 cm | |
Net weight | 13kg | |
Shipping - gross weight | 17kg | |
Performance Specifications | Peak wavelength | 455 to 470 nm |
Lamp duration | 60,000hrs | |
Average spectral irradiance | Top light - 34.8 µW.cm-2.nm-1 | |
Bottom light - 50.4 µW.cm-2.nm-1 | ||
Effective surface area | Top light - 53 cm x 24 cm | |
Bottom light - 48 cm x 20 cm | ||
Irradiance uniformity ratio | Top light - 0.51 (IEC compliant --> 0.4) | |
Bottom light - 0.72 (IEC compliant --> 0.4) | ||
Electrical Specifications | Power characteristics | 60W, 100-240VAC, 47/63Hz |
COTS External PS | International medical safety approvals | |
ANSI/AAMI/EN 60601-1, UL/TUV | ||
Class I construction is standard (ground required) | ||
100k hours MTBF | ||
Energy star efficiency level V compliant | ||
RoHS compliant | ||
Over voltage / over current protection | ||
Environmental Specifications | Operating | Ambient temp: +10°C to +35°C |
Humidity: 0% to 90% RH non condensing | ||
Atmospheric pressure: 70-106kPa | ||
Sealing: IPX3 | ||
Transport and storage | Ambient temp: 0°C to +50°C | |
Humidity: 0% to 90% RH non condensing | ||
Atmospheric pressure: 70-106kPa | ||
Exclusions | Not for use inside an incubator | |
Not for use inside any radiant warmer | ||
Regulatory | EN 60601-1:2006 + A1:2013 | |
EN 60601-1-2:2015 | ||
EN 60601-2-50: 2009 | ||
EN 60601-1-6:2010 | ||
Biocompatibility Tests | ||
ISO: 10993-Part 10 | Skin sensitisation | |
ISO 10993-Part 5 | Cytotoxicity direct contact | |
ISO:10993-Part 10 | Primary skin irritation |
- Home
- About Us
-
Shop
-
Primary Care
- Stethoscope
- Sphygmomanometer
- Blood Pressure Monitor
- Blood Glucose Meter
- Height and Weight
- Pulse Oximeter
- Spirometers
- Examination Table
- X- Ray Film Viewers
- Human Anatomical Models
- DVT pump
- Defibrillator
- General Practice
- Pediatrics
- Emergency medicine
- General Practice
- Electronic Ventilator
- Emergency medicine, Resuscitation
- ICU (Intensive Care Unit)
- Secondary Care
- Accessories
- Ultra sound
- Medical Technical Facilities
- Furniture, Logistics
-
ECG
- Laboratory
- Medical Consumables
- Orthopedic surgery
- Medical Imaging
- Surgery unit
- Wellness
- Gastroenterology
- Cardiology
- Neurology
- Automatic Medical Gas Supply unit
-
Home Use
- EEG
-
Primary Care
- Contact Us
- FAQ's
MTTS Firefly phototherapy
- Home
- MTTS Firefly phototherapy
MTTS Firefly phototherapy
- Rs
The MTTS Firefly is a state-of-the-art phototherapy device designed for effective treatment of neonatal jaundice. Developed in collaboration with Design that Matters (Massachusetts, USA), and East Meets West Foundation (California, USA), the machine features intuitive design, double-sided lighting, compact size, user-friendly control panel, and removable single-infant bassinet. Firefly is designed to treat newborn jaundice in the mother’s room – the best way to support mother-child bonding and breastfeeding.
Clinical evaluations indicate that Firefly’s double-sided design reduces total treatment time by 40% on a typical patient compared to single-sided LED phototherapy, allowing earlier discharge from the hospital, lowering incidence of newborn infection and freeing resources to treat more infants.
Quality
The Firefly was designed through a partnership between MTTS, Design that Matters (Massachusetts, USA), and East Meets West Foundation (California, USA).
Innovation
The Firefly key innovation is to use top and bottom phototherapy with long-life, high power LEDs, providing intensive treatment in a compact package.
Care
The Firefly can be installed directly in the mother’s recovery room, thus promoting mother-child bonding, regular breast-feeding and close maternal supervision.
Key Features
Overall
Compact form fits in an infant cart or mother’s bed.
Handholds facilitate easy portability.
Tight seams to prevent dirt build-up.
Curved surfaces for easy cleaning.
Sealed to keep out dust, and liquids (IPX3).
No moving parts, no internal fans.
Medical-grade power supply (not pictured).
1. Top Light
Fixed at the most clinically effective height.
Slim form for easy patient observation.
Curved top prevents placement of objects that may fall on the infant.
Soft edges to protect infant during placement.
2. Bottom Light
Durable thick-walled plastic.
Tilted bottom directs heat off back edge of device.
3. Bassinet
Removable for patient transport and diagnostics.
Comfortable handhold around the rim.
Clear scratch-resistant plastic for easy observation.
Smooth surface for quick wipe-down.
4. Control Panel
Single power button controls both lights.
Single light intensity setting.
Treatment hours and total device hours display.
Effective
The Firefly provides high-intensity light output in an intuitive package. The toplight is fixed at the most effective height for precise delivery of phototherapy.
Compact
The compact form makes the Firefly fit to be used in the mother’s room, even in space-constrained neonatal care settings, promoting a family-centred healing environment.
Cost Efficient
The Firefly combines low initial cost with long-life LEDs that operate with only 30 watts and are rated to last up to 44.000 hours before requiring replacement.
Technical Specifications
Ordering Information and Servicing
Ordering Information Firefly Unit FF1-1000 Main Unit Bassinet FF1-1010 Gel Mattress FF1-1011 Accessories Battery Backup FF1-10202 2hours - internal Trolley FF1-1040
Eye patch (5 pieces) YPT-1013 Adjustable
Mains Power Cable COR-10001EU Standard
COCOR-10002 US/VN Standard
COR-10003 UK Standard
COR-10004 AUS Standard
COR-10005 India Standard
User Manual FF1- 1030 English
Quick Reference Guide FF1-1031 English
The MTTS Firefly is a state-of-the-art phototherapy device designed for effective treatment of neonatal jaundice. Developed in collaboration with Design that Matters (Massachusetts, USA), and East Meets West Foundation (California, USA), the machine features intuitive design, double-sided lighting, compact size, user-friendly control panel, and removable single-infant bassinet. Firefly is designed to treat newborn jaundice in the mother’s room – the best way to support mother-child bonding and breastfeeding.
Clinical evaluations indicate that Firefly’s double-sided design reduces total treatment time by 40% on a typical patient compared to single-sided LED phototherapy, allowing earlier discharge from the hospital, lowering incidence of newborn infection and freeing resources to treat more infants.
Quality
The Firefly was designed through a partnership between MTTS, Design that Matters (Massachusetts, USA), and East Meets West Foundation (California, USA).
Innovation
The Firefly key innovation is to use top and bottom phototherapy with long-life, high power LEDs, providing intensive treatment in a compact package.
Care
The Firefly can be installed directly in the mother’s recovery room, thus promoting mother-child bonding, regular breast-feeding and close maternal supervision.
Key Features
Overall
Compact form fits in an infant cart or mother’s bed.
Handholds facilitate easy portability.
Tight seams to prevent dirt build-up.
Curved surfaces for easy cleaning.
Sealed to keep out dust, and liquids (IPX3).
No moving parts, no internal fans.
Medical-grade power supply (not pictured).
1. Top Light
Fixed at the most clinically effective height.
Slim form for easy patient observation.
Curved top prevents placement of objects that may fall on the infant.
Soft edges to protect infant during placement.
2. Bottom Light
Durable thick-walled plastic.
Tilted bottom directs heat off back edge of device.
3. Bassinet
Removable for patient transport and diagnostics.
Comfortable handhold around the rim.
Clear scratch-resistant plastic for easy observation.
Smooth surface for quick wipe-down.
4. Control Panel
Single power button controls both lights.
Single light intensity setting.
Treatment hours and total device hours display.
Effective
The Firefly provides high-intensity light output in an intuitive package. The toplight is fixed at the most effective height for precise delivery of phototherapy.
Compact
The compact form makes the Firefly fit to be used in the mother’s room, even in space-constrained neonatal care settings, promoting a family-centred healing environment.
Cost Efficient
The Firefly combines low initial cost with long-life LEDs that operate with only 30 watts and are rated to last up to 44.000 hours before requiring replacement.
Technical Specifications
Dimensions and Weight | Unit dimensions (overall) | 66 cm x 38 cm x 49.5 cm |
Unit dimensions (top light) | 51.6 cm x 12.5 cm x 6.3 cm | |
Unit dimensions (bottom light) | 64.7 cm x 38 cm x 10 cm | |
Shipping (boxed) dimensions | 45 cm x 73 cm x 55 cm | |
Net weight | 13kg | |
Shipping - gross weight | 17kg | |
Performance Specifications | Peak wavelength | 455 to 470 nm |
Lamp duration | 60,000hrs | |
Average spectral irradiance | Top light - 34.8 µW.cm-2.nm-1 | |
Bottom light - 50.4 µW.cm-2.nm-1 | ||
Effective surface area | Top light - 53 cm x 24 cm | |
Bottom light - 48 cm x 20 cm | ||
Irradiance uniformity ratio | Top light - 0.51 (IEC compliant --> 0.4) | |
Bottom light - 0.72 (IEC compliant --> 0.4) | ||
Electrical Specifications | Power characteristics | 60W, 100-240VAC, 47/63Hz |
COTS External PS | International medical safety approvals | |
ANSI/AAMI/EN 60601-1, UL/TUV | ||
Class I construction is standard (ground required) | ||
100k hours MTBF | ||
Energy star efficiency level V compliant | ||
RoHS compliant | ||
Over voltage / over current protection | ||
Environmental Specifications | Operating | Ambient temp: +10°C to +35°C |
Humidity: 0% to 90% RH non condensing | ||
Atmospheric pressure: 70-106kPa | ||
Sealing: IPX3 | ||
Transport and storage | Ambient temp: 0°C to +50°C | |
Humidity: 0% to 90% RH non condensing | ||
Atmospheric pressure: 70-106kPa | ||
Exclusions | Not for use inside an incubator | |
Not for use inside any radiant warmer | ||
Regulatory | EN 60601-1:2006 + A1:2013 | |
EN 60601-1-2:2015 | ||
EN 60601-2-50: 2009 | ||
EN 60601-1-6:2010 | ||
Biocompatibility Tests | ||
ISO: 10993-Part 10 | Skin sensitisation | |
ISO 10993-Part 5 | Cytotoxicity direct contact | |
ISO:10993-Part 10 | Primary skin irritation |
Ordering Information and Servicing
Ordering Information Firefly Unit FF1-1000 Main Unit Bassinet FF1-1010 Gel Mattress FF1-1011 Accessories Battery Backup FF1-10202 2hours - internal Trolley FF1-1040
Eye patch (5 pieces) YPT-1013 Adjustable
Mains Power Cable COR-10001EU Standard
COCOR-10002 US/VN Standard
COR-10003 UK Standard
COR-10004 AUS Standard
COR-10005 India Standard
User Manual FF1- 1030 English
Quick Reference Guide FF1-1031 English
Add a Review
You Might Also Like
Add Related products to weekly line up
In The Same Category
39 other products in the same category:
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New
- New