OSEN 9000 Patient Monitor
Features
- 12.1" color TFT LCD screen, resolution: 800*600
- 5 kinds of interfaces: Standard/Big font/6-leadn ECG/Trend gragh/OxyCRG
- Maximum 8-channel waveforms display, the waveform
- color can be dened by user (7 colors)
- 96-hour storage and review of trend graphic gram and
- table, 400 groups NIBP data, 1800 alarm events Devote to life technology
- ST and arrhythmia analysis, pacemaker analysis, drug dose calculation
- Anti‐debrillation design
- NIBP over pressure protection
- Plug‐in rechargeable battery, AC/DC
Technical Specification
ECG
Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL,V)
Lead Selection: I, II, III, avR, avL, avF, V
Waveform: 2 ch
Lead Mode: 3 Leads ( R, L, F or RA, LA, LL)
Lead Selection: I, II, III,
Waveform: 1 ch
Gain:
- 2.5mm/mV, 5.0mm/mV,
- 10mm/mV,20mm/mV, auto
- HR and Alarm Range: Adult: 15 ~ 300 bpm-Neo/Ped: 15 ~ 350 bpm
Accuracy: ±1% or±1bpm, which is greater
Resolution: 1 bpm
Sensitivity: > 200 (uVp-p) Differential
Input Impedance: > 5 MΩ CMRR :
Monitor > 105 dB
- Operation > 105 dB
- Diagnosis > 85 dB Electrode
Offset Potential: ±300mV Leakage
Current: < 10 uA
Baseline Recovery: < 3 S After Defi.
ECG Signal Range: ±8 m V ( Vp-p )
Bandwidth:
- Surgery: 1 ~ 20 Hz
- Monitor: 0.5 ~ 40 Hz
- Diagnostic: 0.05 ~ 130 Hz
Calibration Signal: 1 (mVp-p), Accuracy:5% ST
Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting:
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, BRADY, MISSED BEATS, PNP, PNC, Alarm Available, Review Available
RESP
Method: Impedance Between R-F(RA-LL)
Differential Input Impedance: > 2.5 MΩ
Measuring Impedance Range: 0.3~3 Ω
Base line Impedance Range: 200~4 KΩ
Bandwidth: 0.1~2.5Hz
Resp. Rate:
Measuring and Alarm Range:
- Adult: 0 ~ 120 Brpm
- Neo/Ped: 0 ~ 150 Brpm
- Resolution: 1 Brpm
- Accuracy: ±2 Brpm
- Apnea Alarm: 10 ~ 40
NIBP
Method: Oscillometric Mode:
Manual, Auto, STAT
Measuring Interval in Auto Mode:
1,2,3,4,5,10,15,30,60,90,120,180,240,480 (Min)
Measuring Period in STAT Mode: 5 Min
Pulse Rate Range: 40 ~240 bpm
Alarm Type: SYS, DIA, MEAN
Measuring and Alarms Range:
Adult Mode
SYS: 40 ~240 mmHg
- ST and arrhythmia analysis, pacemaker analysis, drug does calculation
- Anti- defibrillation design
- NIBP over pressure protection
- Plug-in rechargeable battery. AC/DC avilable
- Support wire or wireless network
DIA: 10 ~ 215 mmHg
MEAN: 20 ~ 235 mmHg
Pediatric Mode
- SYS: 40 ~ 200 mmHg
- DIA: 10 ~ 150 mmHg
- MEAN: 20 ~ 165 mmHg
Neonatal Mode
- SYS: 40 ~ 135 mmHg
- DIA: 10 ~ 100 mmHg
- MEAN: 20 ~ 110 mmHg
Resolution: 1mmHg
Accuracy: Maximum Average Error ±5mmHg
Maximum Standard Deviation ±8mmHg
Overpressure Protection Adult
Mode: 297±3 mmHg
Pediatric Mode: 240±3 mmHg
Neonatal Mode: 147±3 mmHg
SpO2
Measuring Range: 0 ~ 100 %
Alarm Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100% 2 %
0% ~ 69% unspecied
Actualization Interval: about 1 Sec.
Alarm Delay: 10 Sec.
Pulse Rate:
- Measuring and Alarm Range: 20~300 bpm
- Resolution: 1bpm
- Accuracy: ±3bpm
TEMP
Channel: 2
Measuring and Alarm Range: 0 ~ 50 °C
Resolution: 0.1 °C
Accuracy: ±0.2 °C
Actualization Interval: about 1 Sec.
Average Time Constant:<10 sec
IBP
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range:
- ART: 0 ~ 300 mmHg
- PA: -6 ~ 120 mmHg
- CVP/RAP/LAP/ICP: -10 ~ 40 mmHg
- P1 / P2: -10 ~ 300 mmHg
Press Sensor:
Sensitivity: 5 uV/V/mmHg
Impedance: 300-3000Ω
Resolution: 1 mmHg
Accuracy: ±2% or 1mmHg which is greater
Actualization Interval: about 1 secretary
Battery: Rechargeable Li-ion battery, 14.8V 2200mAh
Standard: 3/5- Lead ECG, RESP, Spo2, NIBP,PR, TEMP
Optional: Nellcor SpO2, Mainstream/ Sidestream EtCo2, 1/2 channel IBP, Touch Screen, Thermal Recorder, Wall mount, Trolley, Central Station